Heart Rhythm
Lepu Medical > Product Information > Heart Rhythm > Implantable Pacemaker


2312 Implantable Pacemaker








Product Features



·    9 exact programmable parameters

·    Compatible analyzer and programmer are available

·    Events counter


·    Exact pacing threshold, P/R wave, lead impedance and battery voltage            measurement

·    Run-way and crystal oscillator protection

·    Not deplete pacemaker battery energy when programming and telemetry

·    Longevity: over 10 years

·    Compatible analyzer and programmer are available












Parameters



Physical characteristics

.      Thickness: 6mm

.      Weight: 20.8g

.      Volume: 8.8cm

Parameters

Range

Tolerance

Pacing mode

SSI/SST/SOO

-

Basic rate (ppm)

30…(1)…100…(10)…160

±10ms

Basic Interval (ms)

2000…375

±10ms

Pulse amplitude (V)

0.1…(0.1)…8.1

0.06-0.18, 0.25-0.79, 0.86-1.40,

(0.1-0.9)V±0.1V

(1.0-8.1)V±10%

Pulse width (ms)

1.47-1.59

0.06(step)

(0.060.98)ms±0.1ms

(1.041.59)ms±10%

Sensitivity (mV)

0.5…(0.5)…10

(0.51.5)mV±0.5mV

(2.010.0)mV±30%

Refractory period (ms)

200…(25)…500

±25ms

Hysteresis (ms)

0…(50)…350

±50ms

Pacing Polarity

Unipolar/Bipolar

-

Sensing Polarity

Unipolar/Bipolar

-

 










Ordering Information



Catalogue No.

Function

2312S

1. Interrogate

2. Test battery voltage

3. Pacing event counter

4. Sensing event counter

5. EMI event counter

2312M

1. Interrogate

2. Test battery voltage

3. Pacing event counter

4. Sensing event counter

5. EMI event counter

6. Threshold test

7. P/R Amplitude test

8.Lead impedance test











Contact Us



If you have any query about the product information, please contact our International Sales & Marketing Department: international@lepumedical.com


SOURCE: Lepu Medical International Sales & Marketing Department


  SITE MAP   RECRUITMENT   CONTACT US
POWERED BY LEPU MEDICAL
© Copyright  1999-2014  lepumedical.com,  All rights reserved
版权所有    乐普(北京)医疗器械股份有限公司    京ICP备13012321号