7-inch touch screen of SpO2/ECG waveform and numeric display
Table top design with user-friendly operation
Up to 12000 groups of NIBP, 2000 groups SpO2 and event record
Historic data records can be reviewed in waveform, list or trend graph
Data export and software upgrading can be done by USB cable
Flexible configurations among SpO2, NIBP, TEMP and ECG
Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available
NIBP | Technique | Oscillometric |
Measurement accuracy | Maximum mean difference: ±5 mmHg | |
Maximum standard deviation: 8 mmHg | ||
Measuring range | ||
Systolic pressure | Adult: 40mmHg~275mmHg | |
Pediatric: 40mmHg~200mmHg | ||
Neonate: 40mmHg~135mmHg | ||
Diastolic pressure | Adult: 10mmHg~210mmHg | |
Pediatric: 10mmHg~150mmHg | ||
Neonate: 10mmHg~98mmHg | ||
Mean arterial pressure | Adult: 20mmHg~230mmHg | |
Pediatric: 20mmHg~165mmHg | ||
Neonate: 20mmHg~110mmHg | ||
Measurement mode | Manual, Auto, STAT | |
Overpressure protection limi | Adult: 300mmHg | |
Pediatric: 240mmHg | ||
Neonate: 150mmHg | ||
SpO2 | Transducer | Dual-wave length LED |
SpO2 measuring range | 35%~100% | |
SpO2 measuring accuracy | 3% for range from 70% to 100% | |
Low perfusion performance | As low as 0.4% | |
Pulse Rate | Measuring range | 30bpm~240bpm |
Measuring accuracy | ±2bpm or ±2%, whichever is greater |
Catalogue No. | PC-900plus |