3.5-inch high-resolution LCD display of SpO2/ECG waveform
Tabletop design with user-friendly operation
Up to 12000 groups of NIBP, 3000 groups SpO2 and event record
Historic data records can be reviewed in waveform, list or trend graph
Data export and software upgrading can be done by USB cable
Flexible configurations among SpO2, NIBP, TEMP and ECG
Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available
NIBP | Technique | Oscillometric |
Measurement accuracy | Maximum mean difference: ±5 mmHg | |
Maximum standard deviation: 8 mmHg | ||
Measuring range | ||
Systolic pressure | Adult: 40mmHg~275mmHg | |
Pediatric: 40mmHg~200mmHg | ||
Neonate: 40mmHg~135mmHg | ||
Diastolic pressure | Adult: 10mmHg~210mmHg | |
Pediatric: 10mmHg~150mmHg | ||
Neonate: 10mmHg~95mmHg | ||
Mean arterial pressure | Adult: 20mmHg~230mmHg | |
Pediatric: 20mmHg~165mmHg | ||
Neonate: 20mmHg~110mmHg | ||
Measurement mode | Manual, Auto, STAT | |
Overpressure protection limi | Adult: 300mmHg | |
Pediatric: 240mmHg | ||
Neonate: 150mmHg | ||
SpO2 | Transducer | Dual-wave length LED |
SpO2 measuring range | 35%~100% | |
SpO2 measuring accuracy | 3% for range from 70% to 100% | |
Low perfusion performance | As low as 0.4% | |
Pulse Rate | Measuring range | 30bpm~240bpm |
Measuring accuracy | ±2bpm or ±2%, whichever is greater | |
ECG | Input dynamic range | ±0.5mVp~±5mVp |
HR measuring range | 15bpm~350bpm | |
HR measuring accuracy | ±1% or ±2bpm, whichever is greater | |
HR alarm delay time | ≤10s | |
Sensitivity selection | ×1/4, ×1/2, ×1, ×2 | |
Sweeping speed | 25mm/s | |
ECG noise level | ≤30µVp-p | |
ECG input loop current | ≤0.1µA | |
Differential input impedance | ≥10MΩ | |
Common-mode rejection ratio | ≥105dB | |
TEMP | Measuring range | 21.0℃ ~50.0℃ |
Measuring accuracy | ±0.2℃ for range 25℃ ~45℃ | |
Others | Power supply | 100~240Vac, 50/60Hz |
Built-in battery | 11.1V/4400mAh Li-ion battery | |
Display | 7-inch touch screen | |
Configurations | PC-900SN | SpO2, PR, NIBP monitoring |
PC-900SNT | SpO2, PR, NIBP and TEMP monitoring | |
PC-900SNTE | SpO2, PR, NIBP, TEMP and ECG monitoring | |
Option | Nellcor SpO2, Infrared Ear temp probe, Printer, Clamp, Trolley |
Catalogue No. | PC-900plus |
*Product registration and availability vary by country. For more information on product availability, please contact us here.