country
Contact Us
PC-900pro, Vital Signs Monitor
PC-900pro, Vital Signs Monitor

PC-900pro, Vital Signs Monitor

Features

  • 3.5-inch high-resolution LCD display of SpO2/ECG waveform

  • Tabletop design with user-friendly operation

  • Up to 12000 groups of NIBP, 3000 groups SpO2 and event record

  • Historic data records can be reviewed in waveform, list or trend graph

  • Data export and software upgrading can be done by USB cable

  • Flexible configurations among SpO2, NIBP, TEMP and ECG

  • Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available



Technical Specifications

Ordering Information

NIBPTechniqueOscillometric
Measurement accuracyMaximum mean difference: ±5 mmHg
Maximum standard deviation: 8 mmHg
Measuring range
Systolic pressureAdult: 40mmHg~275mmHg
Pediatric: 40mmHg~200mmHg
Neonate: 40mmHg~135mmHg
Diastolic pressureAdult: 10mmHg~210mmHg
Pediatric: 10mmHg~150mmHg
Neonate: 10mmHg~95mmHg
Mean arterial pressureAdult: 20mmHg~230mmHg
Pediatric: 20mmHg~165mmHg
Neonate: 20mmHg~110mmHg
Measurement modeManual, Auto, STAT
Overpressure protection limiAdult: 300mmHg
Pediatric: 240mmHg
Neonate: 150mmHg
SpO2TransducerDual-wave length LED
SpO2 measuring range35%~100%
SpO2 measuring accuracy3% for range from 70% to 100%
Low perfusion performanceAs low as 0.4%
Pulse RateMeasuring range30bpm~240bpm
Measuring accuracy±2bpm or ±2%, whichever is greater
ECGInput dynamic range±0.5mVp~±5mVp
HR measuring range15bpm~350bpm
HR measuring accuracy±1% or ±2bpm, whichever is greater
HR alarm delay time≤10s
Sensitivity selection×1/4, ×1/2, ×1, ×2
Sweeping speed25mm/s
ECG noise level≤30µVp-p
ECG input loop current≤0.1µA
Differential input impedance≥10MΩ
Common-mode rejection ratio≥105dB
TEMPMeasuring range21.0℃ ~50.0℃
Measuring accuracy±0.2℃ for range 25℃ ~45℃
OthersPower supply100~240Vac, 50/60Hz
Built-in battery11.1V/4400mAh Li-ion battery
Display7-inch touch screen
ConfigurationsPC-900SNSpO2, PR, NIBP monitoring
PC-900SNTSpO2, PR, NIBP and TEMP monitoring
PC-900SNTESpO2, PR, NIBP, TEMP and ECG monitoring
OptionNellcor SpO2, Infrared Ear temp probe, Printer, Clamp, Trolley