5 6 inch color LCD touch screen, 12-lead simultaneously acquisition
Large internal capacity up to 5000 files more ECG files can be stored via USB port
Freeze, Pre-10-second print and Trigger print function to observe any abnormal ECG waveforms
HR detection alarm and pacemaker detection supported
Cine-loop function: saved ECG waveform can be played back
Auto-save function: users can choose whether to automatically save the files with the printout or not
Support magnetic card reader, bar code scanner, make information input more conveniently
Support external laser printer to achieve A4 size report output
Data management software on PC(optional)
Reliable automatic measurement & Interpretation tested with CSE&AHA database
Safety Classification | Type CF |
Safety Standard | MDD93/42/EEC, IEC60601-2-25 IEC60601-1, IEC60601-2-51: 2003, ANSI/AAMI EC-11, EN 1441-1997 |
Input Circuit | Floating, protection against defibrillator and pacemaker detection |
Lead | Standard 12-lead |
Input Impedance | ≥50ΜΩ |
Input Circuit Current | ≤50nA |
Leakage Current | <10μA |
Calibration Voltage | 1mV±2% |
A/D Convertor | 24bits |
Frequency Response | 0.05Hz~165Hz |
Time Constant | ≥3.2s |
Noise Level | <12μVp-p |
Filter | AC: 50/60Hz (-20dB); EMG:25Hz/35Hz/45Hz(-3dB); ADS and Baseline auto-adjustment |
Sensitivity | Auto, 2.5, 5, 10, 20,40 (mm/mV) ±5 % |
Measurement | HR, PR/QT/QTC Interval, P/QRS/T Axis etc. |
Patient Data | Date, time, ID, name, gender, age, height, weight, blood pressure, etc. |
System Data | Sensitivity, Speed, Institution, etc. |
Analysis Report | On/Off selectable |
Printer | High speed and sensitivity thermal dot recording system |
Printing System | High speed, high sensitivity thermal matrix printing system |
Paper Speed | 5/6.25/10/12.5/25/50mm/s (±3%) |
Printing Paper | Roll paper112mm x 20m(30m) |
Power Supply | AC: 100V-240V(±10%), 50/60Hz(±1Hz) |
DC: More than 2-hour continuous operation time with lithium battery after fully charged | |
External Input | 10mm/V ± 5%; Input Impedance=100KO |
ExternalOutput | 1.0V/mV ± 5%; Input Impedance=100O |
EquipmentSize: | 345 x 300 x 80mm |
Carton size | 400 x 180 x 335mm |
Net weight | 2.5Kg(battery included) |
Gross weight | 4.5Kg |
Safety Standard | IEC Class I, type CF |
Acquisition Mode | 12 lead acquisition simultaneously |
Input Circuit | Floating, protection against defibrillator and pacemaker |
A/D Converter | 12 bit |
Time Constant | ≥3.2s |
Frequency Response | 0.05Hz ~ 150Hz(-3dB) |
Sensitivity | 2.5, 5, 10, 20, 40 (mm/mV), Auto |
Input Impedance | ≥50MΩ |
Input Circuit Current | ≤50nA |
Calibration Voltage | 1mV±2% |
Noise Level | <15μVp-p |
Anti Baseline Drift | Automatic |
Filter | EMG Filter: 35Hz ~ 45Hz (-3dB) |
ADS Filter: Auto (0.15-0.50Hz) | |
HUM Filter: 50Hz/60Hz (-20dB) | |
CMRR | >100dB(with AC filter ) |
Patient Current Leakage | <10μA |
Writer Speeds | 5/6.25/10/12.5/25/50mm/s |
Recording Paper | 112mm(width), 20m/30m(length), roll paper |
Power Supply | AC: 110/220V, 50/60Hz |
DC: 14.4V, built-in rechargeable lithium battery | |
Dimension | 345mm×300mm×80mm; |
415mm×195mm×380mm(Carton) | |
Weight | 2.5kgs(N,W) 4.5kgs(G.W) |
Catalogue No. | ECG-1106L |
Catalogue No. | ECG-1106G |