Multi-parameter of Medical Patient Monitor

12.1-inch high resolution touch screen with LED backlight

Arrhythmia analysis and S-T segment measurement

Adult/Pediatric/Neonate measurement modes

Up to 9 waveform displays, 140-hour ECG waveform data storage and recall

2-year warranty on the main unit and half year warranty on accessories



UP-7000 Multi-parameter Medical Patient Monitor Details

UP-7000 Features

• 12.1" high-resolution TFT display with LED backlight

• Arrhythmia analysis and S-T segment measurement

• Protection against defibrillator discharge

• Adult/Pediatric/Neonate measurement modes

• Visual and audible alarms; Networking capability

• Up to 9 waveforms simultaneously display

• 140-hour ECG waveform data storage and recall

• 2000-hour data trends with a graphic and tabular view

• 2000 groups event, ARR and SpO2 storage

• Built-in lithium battery; Touch screen optional 




Technical Specifications

Ordering Information

NIBPTechniqueOscillometric
Typical measuring time<30 seconds (typical adult cuff)
Initial cuff inflation pressureAdult: <175mmHg
Pediatric: <135mmHg
Neonate: <65mmHg
Overpressure protection limitAdult: 300mmHg
Pediatric: 240mmHg
Neonate: 150mmHg
Measuring range
Systolic pressureAdult: 40mmHg~275mmHg
Pediatric: 40mmHg~200mmHg
Neonate: 40mmHg~135mmHg
Diastolic pressureAdult: 10mmHg~210mmHg
Pediatric: 10mmHg~150mmHg
Neonate: 10mmHg~95mmHg
Mean arterial pressureAdult: 20mmHg~230mmHg
Pediatric: 20mmHg~165mmHg
Neonate: 20mmHg~110mmHg
Measurement accuracyMaximum mean difference: ±5 mmHg
Maximum standard deviation: 8 mmHg
Measurement modeManual, Auto, STAT
Automatic measuring intervals1~480min
TEMPMeasuring range21.0°C~50.0°C
Measuring accuracy±0.2°C for range from 25.0°C~45.0°C
SpO2TransducerDual-wave length LED
SpO2 measuring range0%~100%
SpO2 measuring accuracy2% for range from 70% to 100%
Low perfusion performanceAs low as 0.4%
PR measuring range0bpm~250bpm
PR measuring accuracy±2bpm or ±2%, whichever is greater
ECGInput dynamic range±0.5mVp~±5mVp
HR measuring range15bpm~350bpm
HR measuring accuracy±1% or ±2bpm, whichever is greater
HR alarm delay time≤10s
Sensitivity selection×1/4, ×1/2, ×1, ×2, ×4 and Auto
Sweeping speed6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
ECG noise level≤30µVp-p
ECG input loop current≤0.1µA
Differential input impedance≥10Mohm
Common-mode rejection ratio(CMRR)≥105dB (Monitoring mode)
≥89dB (Diagnostic mode)
Time constant≥0.3s (Monitoring mode)
≥3.2s (Diagnostic mode)
RESPRR measuring range0rpm~120rpm
R measuring accuracy±5% or ±2rpm, whichever is greater
OthersPower supply100~240Vac, 50/60Hz
Built-in battery4400mAh Lithium battery
Display12.1 inch (Resolution 800*600)
Alarming modeAudible-visual alarm
Networking portEthernet port
Standard ConfigurationECG, RESP, SpO 2, NIBP, TEMP, PR
Option2-IBP, EtCO, Nellcor Spo, SunTech NIBP, Cardiac Output, Built-in printer, Cerebral State Monitoring, Central monitoring system, Touch screen