Multi-parameter Patient Monitor
7-inch high resolution color display features a shortcut menu and multiple display modes
1.3kg Portable and lightweight design
12000 groups NIBP list and 2000 groups arrhythmia event
60-hour ECG waveform and 1000-hour trend data recall
2-year warranty on the main unit and half year warranty on accessories
• 7" high resolution(800x480) color display
• Portable design with optional carry bag
• Arrhythmia analysis and ST segment analysis
• Protection against defibrillator interference
• Data export via USB cable with software
• 12000 groups NIBP list and 2000 groups arrhythmia event
• 60-hour ECG waveform and 1000-hour trend data recall
• Pitch tone selection and brightness adjustment
• Pacemaker pulse detection
• Central monitoring system optional
NIBP | Technique | Oscillometric method |
Pressure measuring range | 0 mmHg~300mmHg | |
Typical measurement time | <30 seconds (adult cuff) | |
SYS measuring range | Adult: 40mmHg~275mmHg | |
Pediatric: 40mmHg~200mmHg | ||
Neonate: 40mmHg~135mmHg | ||
MAP measuring range | Adult: 20mmHg~230mmHg | |
Pediatric: 20mmHg~165mmHg | ||
Neonate: 20mmHg~110mmHg | ||
DIA measuring range | Adult: 10mmHg~210mmHg | |
Pediatric: 10mmHg~150mmHg | ||
Neonate: 10mmHg~95mmHg | ||
Measuring accuracy | Maximum mean difference: ±5 mmHg | |
Maximum Standard deviation: 8 mmHg | ||
Measuring mode | Manual, Auto, STAT | |
TEMP | Measuring range | 21℃ ~50℃ |
Measuring accuracy | ±0.2℃ for range from 25℃ ~45℃ | |
TEMP responding time | ≤150s | |
SpO2 | Transducer | Dual-wavelength LED |
SpO2 measuring range | 0%~100% | |
SpO2 measuring accuracy | 2% for range from 70%~100% | |
Low perfusion performance | As low as 0.3% | |
PR measuring range | 0bpm~250bpm | |
PR measuring accuracy | ±2bpm or ±2%, whichever is greater | |
ECG | Input dynamic range | ±(0.5mVp~5mVp) |
Differential input impedance | 10M | |
Frequency response | 1~20Hz (Operation) | |
0.67~40Hz (Monitoring) | ||
0.05~150Hz (Diagnostic) | ||
CMRR | 90dB (Diagnostic) | |
105dB (Monitoring & Operation) | ||
Sensitivity selection | ×1/4, ×1/2, ×1, ×2, ×4 an d Auto | |
Sweeping speed | 6.25, 12.5, 25, 50mm/s, tolerance: ±10% | |
HR display range | 15bpm~350bpm | |
HR display accuracy | ±1% or ±2bpm, whichever is greater | |
RESP | Measuring range | 0~120rpm |
Measuring accuracy | ±5% or ±2 rpm, whichever is greater | |
Others | Power supply | AC100V-240V, 50/60Hz, 60VA |
Internal power supply (Lithium battery) DC 14.8V | ||
Display | 7 inch TFT display, 800x480 resolution | |
Alarming | 3 levels audible-visible alarm | |
Networking | Ethernet | |
Standard Configuration | SpO2, PR, NIBP, ECG, RESP, TEMP | |
Option | SunTech NIBP, CMS, Trolley, Wall mount, Clamp |
Catalogue No. | PC-3000 |